|
Additional information provided in d.9., d.10., h.3., h.6.And h.11.The used device and the lens were returned in the blister tray inside the carton.The plunger lock and lens stop were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was retracted to the fill line.No damage was observed to the device.The lens was returned adhered in dried solution to the exterior of the device nozzle tip.One haptic distal tip edge was chipped.One haptic was dried in a bent position.Lens was cleaned with klrp to further evaluate.The anterior optic surface has a small scrape mark in the travel path of a plunger override.The nozzle was removed and cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A non-qualified viscoelastic was used.The lens was returned outside the used device.Optic anterior damage was observed with is typical in appearance to damage from a previous plunger override.The root cause may be related to a failure to follow the ifu.A non-qualified viscoelastic was indicated.The ifu instructs: during device preparation and implantation of the company iol with the company preloaded delivery system, company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.The ifu instructs: take at least 7 seconds to gently fold the iol by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the ¿fill-to¿ line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.The ifu instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome, since the haptic may be trapped and stretched, and/or pinched and sheared by the moving plunger.Plunger override may occur: due to rapid advancement faster than the ifu recommended rate.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Top coat dye stain testing was conducted with acceptable results.The manufacturer internal reference number is: (b)(4).
|