| Model Number |
H74939171100470 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b - pro code (product code): fge, lit.G4 - premarket / 510(k) #: k141521, k141597.
|
| |
|
Event or Problem Description
|
|
It was reported that a balloon rupture occurred.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified subclavian artery.A 10.0 x 40, 75cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon ruptured upon first inflation at 16 atmospheres for 1 minute.The device was removed without any problem, and the procedure was completed with a different device.There were no patient complications as a result of this event.
|
| |
|
Search Alerts/Recalls
|