• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION FREESTYLE AORTIC ROOT BIOPROSTHESIS; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION FREESTYLE AORTIC ROOT BIOPROSTHESIS; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number FR995-23
Medical Device Problem Codes Patient Device Interaction Problem (4001); Intravalvular regurgitation (4068)
Health Effect - Clinical Codes Cardiogenic Shock (2262); Heart Failure/Congestive Heart Failure (4446); Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 10/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that 16 years and 10 months post implant of this 23mm freestyle aortic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was reported as cardiogenic shock and congestive heart failure (chf) which resulted in rehospitalization.No intervention or additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Updated data: b2: outcome attributed to adverse event, b5: event description, h6: coding.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received that reported subsequently, 6 days later, another manufacturers transcatheter valve was implanted valve-in-valve.The reason for intervention was also reported as severe aortic insufficiency.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE AORTIC ROOT BIOPROSTHESIS
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key23523312
Report Number2025587-2025-05901
Device Sequence Number18577382
Product Code LWR
UDI-Device Identifier00681490265904
UDI-Public(01)00681490265904(17)130903
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P970031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/03/2013
Device Model NumberFR995-23
Device Catalogue NumberFR995-23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2025
Supplement Date Received by Manufacturer01/20/2026
Initial Report FDA Received Date11/11/2025
Supplement Report FDA Received Date02/03/2026
Date Device Manufactured10/02/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age78 YR
Patient SexFemale
Patient Weight73 KG
-
-