|
Terumo medical corporation received the reported information.The doctor stated when he shaped the wire the tip feels like it breaks or was broken.The type of procedure was a left heart catheterization.There was no blood loss.Additional information was received on 27oct2025: the field clinical specialist (fcs) was present for the procedure on (b)(6).During the procedure, the doctor commented that the wire tip seemed broken when he went to shape it.There was no harm caused to the patient.The tip did not separate from the wire however it does feel like it was broken.When was shaped the wire instead of the tip replicating a nice curve, it seemed to kink or break at one point.Additional information was received on 31oct2025: the doctor stated that he has noticed this lately with the tip.
|
|
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.No anomaly was found in the manufacturing record and the shipping inspection record.No similar event was found in the past complaint file.Based on the investigation results, no anomaly was found in the manufacturing record and the shipping inspection record.However, since the actual device was not returned and the analysis of it could not be performed, it was not possible to clarify the cause of occurrence.Ashitaka factory is committed to maintaining the quality of the product by performing the following control.In the product assembling process, 100% visual inspection is performed to ensure that there is no deformation in the coil.After the product assembling process, the outer diameter is measured on 100% basis to ensure that there is no anomaly on it.In the shipping inspection, the sliding resistance is measured per each lot to ensure that there is no anomaly in its lubricity.In the shipping inspection, tensile strength is measured on a sampling basis per product code/lot to ensure that there is no anomaly in it.The instructions for use (ifu) of this product indicates as follows: "if any resistance to the runthrough ns or the dilatation catheter is felt during manipulation or if the shape, behaviour or position of the guide wire's tip seems improper, e.G., in a case where the tip is trapped due to vessel spasm or some other cause or the tip is folded as shown in fig.2, stop manipulating the guide wire (and the catheter) and determine the cause carefully by fluoroscopy and take suitable remedial actions.Then remove the guide wire slowly, without turning, to exchange it for a new one.Continuing guide wire or catheter manipulation in the said situations may result in damage to the vessel, damage to or separation of the guide wire's tip, and/or damage to the dilatation catheter.".
|