Model Number IA-LH-50-10
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Unspecified Infection (1930); Inadequate Osseointegration (2646); Implant Pain (4561)
Date of Event 10/27/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Keystone dental inc.Provides labeling with all implants.Failure to osseointegrate and loss of osseointegration (implant mobility) is identified as an adverse reaction and identified in the list of warnings in all endosseous dental implant labeling.The specific cause for a particular complaint is often not readily identified as there are various factors that contribute to the risk of implant failure.These include patient factors such as prior oral infection, poor bone quality or quantity, systemic conditions such as diabetes and uncontrolled hypertension.Technique factors such as overheating of the implant site may cause necrosis of the bone, preventing growth (pelayo et al., 2008; turkyilmaz & al., 2008).Patient habits such as tobacco use (heitz-mayfield & huynh-ba, 2009), alcohol or drug abuse, poor oral hygiene, and bruxism (salvi & braegger, 2009) may also lead to implant failure.In addition, improper surgical technique can lead to implant failure and/or loss of supporting bone (salvi & braegger, 2009).Additional health effect clinical code=2651.
 
Event or Problem Description
Loss of osseointegration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRINX
Common Device Name
TRNX IMP D50 X 10.5MM LNG
Manufacturer (Section D)
KEYSTONE DENTAL, INC.
13645 alton pkwy ste a
irvine CA 92618
Manufacturer (Section G)
KEYSTONE DENTAL, INC.
13645 alton pkwy ste a
irvine CA 92618
Manufacturer Contact
kathleen labrie
13645 alton pkwy ste a
irvine, CA 92618
9784308257
MDR Report Key23524630
Report Number3005990499-2025-72206
Device Sequence Number10167426
Product Code DZE
UDI-Device IdentifierD768IALH50100
UDI-PublicD768IALH50100
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Dentist
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberIA-LH-50-10
Device Catalogue NumberIA-LH-50-10
Device Lot Number26408
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2025
Initial Report FDA Received Date11/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/09/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexFemale
-
-