• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. COMBISET ACCESS FLOW REVERSE CON TWISTER; ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. COMBISET ACCESS FLOW REVERSE CON TWISTER; ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS Back to Search Results
Catalog Number 03-2794-0
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 11/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
A user facility¿s registered nurse (rn) reported that a fresenius combiset bloodline disconnected at the twister connector one and a half hours after the initiation of a patient¿s hemodialysis (hd) treatment.The machine, a fresenius 2008t machine, alarmed appropriately.Fresenius bloodlines were also in use.There were no issues noted during priming.There were no changes in pressure of blood flow rate during the treatment.The patient¿s estimated blood loss (ebl) was approximately 100ml.There was no patient serious injury or medical intervention required as a result of this event.The patient opted to end treatment two and a half hours early.The patient did not have any issues as a result of ending treatment early.The complaint device is not available to be returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
A user facility¿s registered nurse (rn) reported that a fresenius combiset bloodline disconnected at the twister connector one and a half hours after the initiation of a patient¿s hemodialysis (hd) treatment.The machine, a fresenius 2008t machine, alarmed appropriately.Fresenius bloodlines were also in use.There were no issues noted during priming.There were no changes in pressure of blood flow rate during the treatment.The patient¿s estimated blood loss (ebl) was approximately 100ml.There was no patient serious injury or medical intervention required as a result of this event.The patient opted to end treatment two and a half hours early.The patient did not have any issues as a result of ending treatment early.The complaint device is not available to be returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMBISET ACCESS FLOW REVERSE CON TWISTER
Common Device Name
ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key23525141
Report Number0008030665-2025-02716
Device Sequence Number13312670
Product Code KOC
UDI-Device Identifier00840861100316
UDI-Public00840861100316
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K022536
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03-2794-0
Device Lot Number25ER01104
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2025
Supplement Date Received by Manufacturer02/13/2026
Initial Report FDA Received Date11/11/2025
Supplement Report FDA Received Date02/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS DIALYZER; FRESENIUS DIALYZER; FRESENIUS HEMODIALYSIS MACHINE; FRESENIUS HEMODIALYSIS MACHINE
Patient Age65 YR
Patient SexMale
Patient Weight88 KG
-
-