| Catalog Number |
190766 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
A user facility's biomedical technician (bmt) reported that a 2008t hemodialysis (hd) machine's valve 103 started smoking and cracked open during chemical disinfection.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The valve was the original fresenius part on the machine.The bmt reported that there was no melting, smoke, spark, flame, or arcing observed.The bmt reported that the machine has not had any past problems with failing the electrical leakage test and that there was no damage observed on any other components.The machine has approximately 40,000 hours and is plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet.The bmt replaced the bibag module, which resolved the issue.The bmt reported that the machine has been returned to service without issue.The complaint sample is not available to be returned to the manufacturer for physical evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation and an on-site evaluation was not performed by a fresenius field service technician (fst).The manufacturer was able to determine a causal relationship between the objective evidence provided by the customer and the reported event.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The reported event has been confirmed.
|
| |
|
Event or Problem Description
|
|
A user facility's biomedical technician (bmt) reported that a 2008t hemodialysis (hd) machine's valve 103 started smoking and cracked open during chemical disinfection.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The valve was the original fresenius part on the machine.The bmt reported that there was no melting, smoke, spark, flame, or arcing observed.The bmt reported that the machine has not had any past problems with failing the electrical leakage test and that there was no damage observed on any other components.The machine has approximately 40,000 hours and is plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet.The bmt replaced the bibag module, which resolved the issue.The bmt reported that the machine has been returned to service without issue.The complaint sample is not available to be returned to the manufacturer for physical evaluation.
|
| |
|