WILSON-COOK MEDICAL INC 6 SHOOTER SAEED MULTI-BAND LIGATOR; LIGATOR, ESOPHAGEAL Back to Search Results
Model Number G22553
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
During preparation for an unknown endoscopic procedure, the physician selected a cook 6 shooter saeed multi-band ligator.It was reported that the trigger cord was broken while nurse was loading the barrel to the tip of the endoscope.They opened another box of mbl-6 to finish procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Continued: section e phone: (b)(6).Section e country: hong kong.Investigation evaluation: a product evaluation was performed only by the photo provided in response to this report because the product said to be involved was not provided to cook for evaluation.Per the photo provided we cannot complete a full evaluation.Without the product or substantial evidence to contradict the complaint, it is considered confirmed based solely on statements and the photo describing the event.The photo shows the trigger cord with 4 bands tangled on it; the trigger cord does not appear to be broken, as the user alleged.The bands were most likely pushed off the barrel onto the trigger cord when the device was being loaded onto the endoscope.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the photo provided confirmed premature deployment of bands.However, we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.A possible contributing factor to premature band deployment includes allowing the trigger cord to become lodged between the barrel and the distal end of the endoscope.This can restrict trigger cord movement and result in premature band deployment if enough tension is applied to pull the cord free.Premature band deployment can also occur if the handle is placed in the firing position before the endoscope is in place inside the patient or even while winding the trigger cord.The user is advised not to apply tension to the trigger cord while winding it on the handle to avoid premature deployment of bands.The instructions for use caution the user: ¿with handle in two-way position, slowly rotate handle clockwise to wind trigger cord onto handle spool until it is taut.Note: care must be taken to avoid deploying a band while winding trigger cord." prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6 SHOOTER SAEED MULTI-BAND LIGATOR
Common Device Name
LIGATOR, ESOPHAGEAL
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key23526744
Report Number1037905-2025-00757
Device Sequence Number10202370
Product Code MND
UDI-Device Identifier00827002225534
UDI-Public(01)00827002225534(17)260408(10)W4928854
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/08/2026
Device Model NumberG22553
Device Catalogue NumberMBL-6
Device Lot NumberW4928854
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2025
Initial Report FDA Received Date11/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENDOSCOPE-OLYMPUS-HQ190.
Patient SexUnknown
-
-