|
The device was returned to olympus for inspection, and the e102 error was not confirmed, the lamp light intensity reduction was confirmed.Updated fields: d8, g3, g6, h2, h3, h6 based on the results of the investigation, for the e102 error, since the device malfunction was not confirmed during evaluation, the definitive root cause of the reported issue could not be determined.For the lamp light intensity reduction, it is likely the following led to the malfunction: stress problem.Olympus will continue to monitor field performance for this device.
|