• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER AND PAYKEL HEALTHCARE; NASAL CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER AND PAYKEL HEALTHCARE; NASAL CANNULA Back to Search Results
Model Number OJR416
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
A distributor in japan reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that the tubing of an ojr416 optiflow junior 2 nasal cannula was detached from the cannula.There were no reported patient involvement.
 
Additional Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Additional Manufacturer Narrative
(b)(4).Method: the subject device, ojr416 optiflow junior 2 nasal cannula was returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the evaluation of the returned device, information and photography provided by the distributor and our knowledge of the product.Results: evaluation of the subject device confirmed that the tubing was detached from the cannula end.Conclusion: based on the evaluation of the returned device, information provided and our knowledge of the product, it is most likely that the reported event resulted from the subject device being subjected to excessive force.As part of the manufacturing process, all optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails specifications is discarded.In addition, representative samples from each batch are functionally tested to assess retention strength between the assembled components.If there are any failures the entire batch is quarantined for further investigation.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 nasal cannula / optiflow junior nasal cannula.They also state the following: appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death).Do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).Ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.
 
Event or Problem Description
A distributor in japan reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that the tubing of an ojr416 optiflow junior 2 nasal cannula was detached from the cannula.There were no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER AND PAYKEL HEALTHCARE
Common Device Name
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key23527092
Report Number9611451-2025-01340
Device Sequence Number15738498
Product Code BTT
UDI-Device Identifier09420012435958
UDI-Public(01)09420012435958(10)2103462439(11)241112(17)291112
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K222197
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberOJR416
Device Catalogue NumberOJR416
Device Lot Number2103462439
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2025
Supplement Date Received by Manufacturer01/28/2026
Initial Report FDA Received Date11/11/2025
Supplement Report FDA Received Date02/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-