• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number MF-410295
Medical Device Problem Codes Material Rupture (1546); No Apparent Adverse Event (3189)
Health Effect - Clinical Codes Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported "rupture" and "exchange of textured devices due to patient's concern with product".This record is for the right side.Device remains implanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, d6b, h.6.
 
Event or Problem Description
Healthcare professional reported "rupture" and "exchange of textured devices due to patient's concern with product".Healthcare professional later reported "there was no rupture".This record is for the right side.Device has been explanted and replaced.The device will be returned.
 
Additional Manufacturer Narrative
Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: anxiety - product/ procedure: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (opening ) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Event or Problem Description
Healthcare professional reported "rupture" and "exchange of textured devices due to patient's concern with product".Healthcare professional later reported "there was no rupture".This record is for the right side.Device has been explanted.
 
Additional Manufacturer Narrative
Additional and/or changed data: b2 b5 h6.This record is no longer reportable to the fda and will be un-reported.
 
Event or Problem Description
This record is no longer reportable to the fda and will be un-reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key23528846
Report Number9617229-2025-19529
Device Sequence Number21087912
Product Code FTR
UDI-Device Identifier10888628003453
UDI-Public10888628003453
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/07/2019
Device Catalogue NumberMF-410295
Device Lot Number2632001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/27/2025
Supplement Date Received by Manufacturer12/10/2025
01/29/2026
02/03/2026
Initial Report FDA Received Date11/11/2025
Supplement Report FDA Received Date12/27/2025
02/03/2026
02/04/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexFemale
Patient Weight56 KG
-
-