Model Number 406849
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One 8.5f fast-cath introducer sheath and dilator were received for evaluation.The dilator distal tip had been split, and the distal tip of the sheath had been punctured.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the device damage remains unknown.
 
Event or Problem Description
This report is to advise of a puncture found in the tip of the sheath during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.
Common Device Name
Introducer, catheter
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
wendy mccarty
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key23529430
Report Number3005334138-2025-00687
Device Sequence Number21378987
Product Code DYB
UDI-Device Identifier05414734203388
UDI-Public(01)05414734203388(10)10574421(17)270930
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number406849
Device Catalogue Number406849
Device Lot Number10574421
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2025
Initial Report FDA Received Date11/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-