| Model Number |
DDMC3D4 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
10/22/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient received possibly inappropriate anti-tachycardia pacing (atp) and high-voltage therapy for supra ventricular tachycardia (svt) that the implantable cardioverter defibrillator (icd) classified as ventricular tachycardia (vt).It was also noted that the right atrial (ra) lead exhibited intermittent undersensing noted on a stored episode resulting in the misclassification of svt versus a vt episode.The patient received atp therapy for this episode.The icd and lead remain in use.No further patient complications have been reported as a result of this event.
|
| |
|