|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The curved tip stapler instrument was analyzed and found to have a broken grip cable at the pivot tube.The cable was fully broken.The segment of the cable that contains the crimp was still intact upon initial inspection.The cable was tugged on to confirm the breakage, and was separated from the instrument during the inspection.There was no evidence of discoloration or corrosion/contamination on the cables to indicate a reprocessing induced issue.The components surrounding the broken cable did not exhibit abnormal sharp edges or damage that would have contributed to the cable breaking.Additional observation(s) related to customer reported complaint: the instrument was found to have two firing failures based on log review, which were not replicated through in-house testing.The stapler was placed on an in-house system and was unable to open.The broken grip cable did not allow operation of the jaws.Therefore, the stapler could not be tested for its firing capabilities.Common causes of broken - distal instrument grip cables, whether partial or full, are commonly attributed to damage to the cable through external collisions, during transport/storage, during use, or during reprocessing.The root cause of the failed fires is attributed to a broken grip cable.Blank mdr fields: the missing patient information in sections a and b was either unknown, unavailable, not provided, or not applicable.The expiration date for section d4 is not applicable.Field d6 is blank because the product is not implantable.Information for the blank fields in section e1 is not available.Field e4 is blank because it is unknown if the initial reporter submitted a report to the fda.Fields g5 and g7 are not applicable.Field h10 is blank as there are no related report numbers.
|