Model Number FX2100X15B
Medical Device Problem Codes Circuit Failure (1089); Mechanical Problem (1384); Failure to Calibrate (2440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a 3-way failure occurred on a trilogy evo.No medical intervention was specified.The device was not reported to be in clinical use at the time of the allegation.The trilogy evo was returned to the third-party service center for evaluation.During the evaluation it was noted that a failure of the device's circuit board and obm subassembly was discovered in the error log.In conclusion, the device's system circuit board and obm subassembly was replaced to address the issue.The device was tested and passed all final testing.A final report is being submitted.If new information becomes available, a follow-up/supplemental report will be completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23531894
Report Number2518422-2025-111728
Device Sequence Number18397394
Product Code CBK
UDI-Device Identifier00606959061019
UDI-Public(01)00606959061019
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberFX2100X15B
Device Catalogue NumberFX2100X15B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2025
Initial Report FDA Received Date11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/06/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-