| Model Number |
8888145014 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, prior to dialysis, a crack was observed at the blue (venous) luer connector, and it was also stated that there was a leak.The catheter was repaired.There was no instrument used to loosen or tighten the device.Iodophor was the cleaning agent used on the device.Tego was not utilized.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: mozarc medical conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted one catheter, with a crack on the threads of its venous luer adapter.It was reported that the luer adapter was cracked.The reported issue was confirmed.The most likely cause could not be established from the information available.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all mozarc medical quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a3a, d9, g3, h3, h6 h3 evaluation summary: med conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted one catheter, with a crack on the threads of its venous luer adapter.It was reported that the luer adapter was cracked.The reported issue was confirmed.The most likely cause could not be established from the information available.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|