• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OPTICROSS? 6; TRANSDUCER, ULTRASONIC, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OPTICROSS? 6; TRANSDUCER, ULTRASONIC, DIAGNOSTIC Back to Search Results
Model Number H7495181360
Medical Device Problem Codes Entrapment of Device (1212); Material Deformation (2976); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a catheter issue occurred and the vessel was occluded.The 90% stenosed target lesion as located in the moderately tortuous and mildly calcified left anterior descending artery (lad).The opticross 6 imaging catheter was advanced for ultrasound examination of the target lesion.Following stenting, the opticross was advanced again, but was difficult to remove and vessel occlusion was noted.The opticross and guidewire were removed together, and the wire broke in the mid-section.The vessel was ballooned and the procedure continued and the lad was successfully opened.The patient fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS? 6
Common Device Name
TRANSDUCER, ULTRASONIC, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key23533475
Report Number2124215-2025-82149
Device Sequence Number18426992
Product Code OBJ
UDI-Device Identifier00191506035433
UDI-Public(01)00191506035433(17)270808(10)37166822
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
K213593
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH7495181360
Device Catalogue NumberH7495181360
Device Lot Number0037166822
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2025
Initial Report FDA Received Date11/12/2025
Date Device Manufactured08/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
Patient Weight41 KG
-
-