| Catalog Number |
0165L20 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that immediately after started using the foley catheter had leakage of sterile water.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed, manufacturing related.The reported failure was able to be reproduced.The product was used for urological care.Based on the evaluation, a tear was found on the rubberized layer causing interlumen leakage.Hence, this complaint was confirmed related to manufacturing process.A potential root cause for this failure could be insufficient latex strength, low/non-uniform rubberize thickness, incorrect wire pressing technique, incorrect stripping technique etc.The labeling and instructions for use were found to be adequate.A review of the manufacturing process indicates that the process is capable of producing of this type of defect.However, current in-process controls per pfmea are adequate to identify the defect or verify the product integrity.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that immediately after started using the foley catheter had leakage of sterile water.
|
| |
|
Event or Problem Description
|
|
It was reported that immediately after started using the foley catheter had leakage of sterile water.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed ¿ manufacturing related.The reported failure was able to be reproduced.The product was used for urological care.Based on the evaluation, a tear was found on the rubberized layer causing interlumen leakage.Hence, this complaint was confirmed related to manufacturing process.A potential root cause for this failure could be insufficient latex strength, low/non-uniform rubberize thickness, incorrect wire pressing technique, incorrect stripping technique etc.The labeling and instructions for use were found to be adequate.A review of the manufacturing process indicates that the process is capable of producing of this type of defect.However, current in-process controls per pfmea are adequate to identify the defect or verify the product integrity.Corrections: d, e, f, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|