ZIMMER BIOMET, INC. COMPREHENSIVE REVERSE SHOULDER; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number 010000589
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Aseptic loosening (4883)
Date of Event 10/17/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: associated product information, part (lot): 00-4349-012-13 (66927809).00-4350-036-00 (66773441).115395 (67171828).180551 (67184547).180554 (67246578).115313 (j7911611).G2: foreign - the event occurred in japan.The customer has indicated that the product is in the process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient experienced pain and implant loosening approximately five (5) weeks after an initial reverse shoulder arthroplasty procedure.A revision procedure was performed in which the baseplate, adapter, poly liner, and glenosphere head were replaced.The surgical technique specified for the device was used during the initial surgery.The revision surgery was successful, and no additional complications have been reported.Attempts have been made, and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; d9; d10; g1; g3; g6; h1; h2; h6.Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products in the d10 narrative below.D10: concomitant medical products, part (lot); mfr report; 115313 (j7911611); 0001825034-2026-00601.Associated product information, part (lot): 00-4349-012-13 (66927809), 00-4350-036-00 (66773441), 115395 (67171828), 180551 (67184547), 180554 (67246578).The product has been received by zimmer biomet and the investigation is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient experienced pain and implant loosening approximately five (5) weeks after an initial reverse shoulder arthroplasty procedure.The glenosphere head was found to have come loose.A revision procedure was performed in which the baseplate, adapter, poly liner, and glenosphere head were replaced.The surgical technique specified for the device was reported to have been used during the initial surgery.The revision surgery was reported as successful with no additional complications.Upon return of the product, the taper adapter was found lodged in the glenosphere.It was reported that no further information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6.The complaint sample was evaluated, and the reported event was confirmed.The taper adapter is angled within the glenosphere and is lodged in place.Visual examination of the returned product identified that the taper adapter returned shows signs of use as well as wear and tear.There is galling on the side of the taper adapter, as shown in the photos.There are also scratches on the visible surface of the taper adapter as well.Measured post height and confirmed etch to confirm the device is conforming.The glenosphere returned also shows signs of use.There are several scratches on the articulating surface.Measured the outer diameter of the device and verified the etch to confirm the glenosphere is conforming to the print specification.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE REVERSE SHOULDER
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23533657
Report Number0001825034-2025-03576
Device Sequence Number9020966
Product Code PHX
UDI-Device Identifier00880304532465
UDI-Public(01)00880304532465(17)350709(10)67218480
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K120121
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number010000589
Device Lot Number67218480
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2025
Supplement Date Received by Manufacturer02/17/2026
04/24/2026
Initial Report FDA Received Date11/12/2025
Supplement Report FDA Received Date03/16/2026
04/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/09/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
H11.; H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-