| Model Number |
AU681-10E CHEMISTRY ANALYZER |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Beckman coulter field service engineer (fse) confirmed that the probable cause was identified as malfunctioning solenoid valves.The fse replaced the solenoid valve to resolve the reported issue.Section a2, a4 and a5: information not provided by the customer.The beckman coulter internal identifier is (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported recovering low glucose (glu) and ion selective electrode (ise) results, which included sodium (na) when run on their au680 chemistry analyzer (pn b96696, sn (b)(6)).There was no report of change to treatment, injury or death associated with this event.The customer did not provide patient data or patient demographics.
|
| |
|