MAUDE Adverse Event Report: CAREFUSION 303, INC. BD PYXIS¿ MEDSTATION¿ ES; AUTOMATED DISPENSING CABINET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 303, INC. BD PYXIS¿ MEDSTATION¿ ES; AUTOMATED DISPENSING CABINET Back to Search Results
Model Number 500001002500
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 06-jun-2023 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that main drawer 3.1 and 3.2 not detected on bus.The field service engineer (fse) replaced the main drawer 3.1 drawer controller card.Functionality was successfully tested using the hardware test application.The system functioned as intended after the field service engineer resolved the issue.
 
Event or Problem Description
It was reported that when using the bd pyxis¿ medstation¿ es had a drawer not detected on bus.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PYXIS¿ MEDSTATION¿ ES
Common Device Name
AUTOMATED DISPENSING CABINET
Manufacturer (Section D)
CAREFUSION 303, INC.
10020 pacific mesa blvd.
san diego CA 92130
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A DE C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX   22250
Manufacturer Contact
brett wilko
10020 pacific mesa blvd.
san diego, CA 92130
MDR Report Key23534408
Report Number2016493-2025-134039
Device Sequence Number13308583
Product Code BRY
UDI-Device Identifier10885403512667
UDI-Public(01)10885403512667
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number500001002500
Device Catalogue Number323
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2025
Initial Report FDA Received Date11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-