| Catalog Number |
08496609190 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Non Reproducible Results (4029)
|
| Health Effect - Clinical Codes |
No Clinical Signs, Symptoms or Conditions (4582); Hyperthyroidism (4924)
|
| Date of Event |
10/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received questionable high results for an unspecified number of patient samples tested with elecsys anti-tshr.The doctors complain that the values are too high as they do not match the clinical status of the patients or the previous values from the patients.The customer provided specific data for six patient samples.No units of measure were provided.The first sample resulted in an anti-tshr value of 2.21.The second sample resulted in an anti-tshr value of 1.87.The third sample resulted in an anti-tshr value of 1.96.The fourth sample resulted in an anti-tshr value of 1.85.The fifth sample resulted in an anti-tshr value of 2.43.The sixth sample resulted in an anti-tshr value of 2.39.
|
| |
|
Additional Manufacturer Narrative
|
|
The serial number of the e411 analyzer is (b)(6).The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Issues have been identified with the elecsys anti-tshr assay, particularly around the 1.75 iu/l medical decision point.Roche diagnostics has issued an urgent medical device correction for this issue, entitled "elecsys® anti-tshr assay ¿ deviations on cobas® e 411, 402, and 801 analyzers." this correction has been reported to fda.Roche diagnostics investigated and confirmed the customers' allegations.The root cause for the variations in the elecsys anti-tshr assay was an insufficiency of the assay's routine production process.For the platform-to-platform deviation (cobas e 411 vs.Cobas e 402/e 801), differences arose due to an insufficient alignment concept between the platforms.These variations were further influenced by updates to the routine production process for this assay.For the lot-to-lot deviation on e 411, measures taken to enhance robustness led to additional shifts.Customers were advised to immediately discontinue using and discard the affected lot numbers: 840183 and 874011 on the cobas e402/e801 analyzer.840177 on the cobas e411 analyzer.Customers were advised to switch to the new unaffected lot numbers: 908953 on the cobas e402/e801 analyzer.906872 on the cobas e411 analyzer.The cobas e601 and e602 modules are not affected and can be used with all currently available lots and with the upcoming lots without restrictions.Medwatch fields h7 and h9 have been updated.
|
| |
|
Search Alerts/Recalls
|