This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h2, h3, h4, h6, h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Review of complaint history identified additional similar complaints for the reported items and no additional complaints for the reported part and lot combinations.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A radiograph was provided and reviewed by a health care professional.Review of the available records identified the following: postoperative ap x-ray of the right hip shows noncemented total hip arthroplasty components in place.The tip of the femoral stem is excluded.Otherwise the femoral component is unremarkable.The acetabular cup is associated with 2 screws superiorly.The more medial of the screws has passed through the screw hole of the acetabular cup and as advanced too far, and is no longer associated with the acetabular cup.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|