MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number ALLURA XPER FD20/20
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to philips that renal biopsy procedure the table¿s longitudinal movement did not perform as intended which delayed an examination.The report indicated that an injury occurred; however, no further details regarding the alleged injury have been provided at this time.An investigation into this complaint has been initiated by philips.
 
Additional Manufacturer Narrative
Philips investigated the reported complaint.Based on the information obtained, the issue occurred prior to patient preparation and entering the examination room.The renal biopsy procedure was rescheduled and later completed successfully, after which the patient was discharged.A philips field service engineer (fse) was dispatched to the customer site and confirmed the reported issue.Review of the system log files showed an ¿application error: ethercat post error,¿ indicating a malfunction of the amc ecat drive.The fse replaced the amc ecat drive, which resolved the issue.The system was repaired, tested, and returned to clinical use in good working condition.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
marjan vos
veenpluis 6
best 5684 -PC
NL   5684 PC
MDR Report Key23535289
Report Number3003768277-2025-014506
Device Sequence Number10187150
Product Code OWB
UDI-Device Identifier00884838054226
UDI-Public(01)00884838054226
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K162859
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberALLURA XPER FD20/20
Device Catalogue Number722038
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2025
Supplement Date Received by Manufacturer01/21/2026
Initial Report FDA Received Date11/12/2025
Supplement Report FDA Received Date02/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-