The gore® cardioform septal occluder instructions for use list pericardial tamponade and significant pleural or pericardial effusion requiring drainage as potential device and procedure related adverse events.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
It was reported the physician selected a 25mm gore® cardioform septal occluder to treat a patent foramen ovale.During deployment the occluder formed improperly and it appeared to be deployed in the tunnel.The physician reloaded the occluder, recrossed the septum, then deployed and locked the device.Upon further assessment, it was noted that the occluder was too small as it didn't capture the whole septum.The occluder was removed without issue; however, following removal a pericardial effusion was noted on intracardiac echocardiography.Pericardiocentesis was performed and the patient was stable.The procedure was concluded without defect closure.
|