| Catalog Number |
33680022 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646)
|
| Date of Event |
06/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Subject: (b)(6).Infinity¿ with adaptis¿ technology total ankle replacement follow-up (itar2).On (b)(6) 2022: patient reports pain in left ankle.Patient has some tenderness sub fibular impingement area.Ct prophecy protocol done to ensure that implant is not loose.On (b)(6) 2023: surgical intervention involving explant of device due to aseptic loosening.Patient discharged (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
Subject: (b)(6).Infinity¿ with adaptis¿ technology total ankle replacement follow-up (itar2).(b)(6) 2022: patient reports pain in left ankle.Patient has some tenderness sub fibular impingement area.Ct prophecy protocol done to ensure that implant is not loose.(b)(6) 2023: surgical intervention involving explant of device due to aseptic loosening.Patient discharged on (b)(6) 2023.
|
| |
|
Search Alerts/Recalls
|