• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC FEMORAL PS CEM LEFT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC FEMORAL PS CEM LEFT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-010-01-0230
Medical Device Problem Codes Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Cyst(s) (1800); Pain (1994); Scar Tissue (2060); Synovitis (2094); Osteolysis (2377); Fibrosis (3167); Joint Laxity (4526); Aseptic loosening (4883)
Date of Event 04/02/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: logic tibia trap tray cem: the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported via legal documentation that approximately 141 months after a left total knee replacement procedure, the patient underwent a revision procedure for pain, aseptic loosening of the patella component, and marked osteolysis from polyethylene wear.Additional information received from the operative report noted that the patient presented with a painful and unstable left knee replacement.The patient was noted to have aseptic loosening of the patella component and marked osteolysis from polyethylene wear.The patient has failed conservative management including activity modification and anti-inflammatory medications.The patient decided to proceed with a revision left total knee replacement.A partial fat pad excision was performed.The patella was subluxated laterally.There was very extensive synovitis and a very thorough synovectomy of the joint was performed.The polyethylene liner was removed.There was noted to be wear of the liner.Next, the femoral component was removed.There was significant fibrous tissue beneath it with osteolysis especially anteriorly.There was no evidence of infection grossly although multiple specimens were sent to pathology.Attention was then turned to the tibia.The prosthesis was removed.Cement was removed.Attention was then directed to the femur.A lamina spreader was inserted laterally and the medial posterior synovitis and scar tissue as well as any posterior osteophytes were then removed.The lamina spreader was switched to the medial side and then the lateral posterior synovitis and scar tissue as well as posterior osteophytes were removed.The remaining intracondylar notch tissue was removed with electrocautery.Attention was then turned to the patella.The patella was noted to be loose and there was some wear on the patella.A freehand patellar cut was made without difficulty and the old patella component was removed.There was some osteolysis of the remaining patella that was debrided.All trial implants were removed.The posterior capsule was cauterized to minimize postoperative bleeding.A periarticular pain injection was injected into the posterior capsule medially and periarticular tissues.Bone ends were irrigated and dried.Bone grafting of small femoral and tibial cysts as well as the intramedullary canal was performed.The knee was copiously irrigated, suctioned and dried thoroughly to prepare for cementation.A competitor tibial cone was impacted into the proximal tibia.Implants were assembled and then cemented into place using cement with antibiotics.The knee was then re-reduced further pressurizing the cement.All excess cement was removed.The cement was allowed to polymerize.The patient was closed and transferred to the recovery room in stable condition.Additional information received from the facility noted that the patient was revised due to femoral knee debonding/loosening.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: g.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: h6, h11 the following sections were corrected: h6 the reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.It was discovered that a bag used in the packaging of our polyethylene inserts had been produced outside of our quality specifications.The use of these non-conforming bags may enable increased oxygen diffusion to the uhmwpe (ultra-high molecular weight polyethylene) insert, resulting in increased oxidation of the material relative to inserts packaged with bags that conform to specifications.For inserts with greater than five years of shelf life, increased oxidation of the inserts can lead to premature polyethylene wear resulting in revision surgery.The non-conforming bags were also used in the packaging of our patella components.An additional hhe and crc were held to assess the risk of the non-conforming packaging specifically related to patella components.Patella components packaged in non-evoh vacuum bags, which have a maximum 5-year shelf-life prior to expiration, have reduced risk of oxidative degradation.Since manufacturing information was not available for the patella component, it could not be confirmed whether the device was packaged in a non-conforming vacuum bag.Additionally, these devices appear to have been implanted over ten years prior to the reported event.Based on the age of the patient¿s devices and the available information, there is no evidence or information to suggest that the reported loosening is manufacturing-related.A review of complaint history determined that no further action is required as no trends were identified.A definitive root cause was unable to be determined; however, may have resulted from prosthesis wear, instability, patellar loosening, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted over ten years prior to the reported event.Potential contributions of users and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM LEFT SZ 3
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key23536179
Report Number1038671-2025-03327
Device Sequence Number18692561
Product Code JWH
UDI-Device Identifier10885862001146
UDI-Public10885862001146
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number02-010-01-0230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2025
Supplement Date Received by Manufacturer02/12/2026
04/16/2026
Initial Report FDA Received Date11/12/2025
Supplement Report FDA Received Date02/12/2026
04/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-