Model Number NAV4680003
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis: product id# (b)(4); lot# em23a032 visual and optical examination identified.It appears that part of the tip of the instrument has been broken off.The metal deformation gives indication that the failure was the result of overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having spinal therapy.It was reported that the tip of the inserter broke when loading the trial on the inserter.No patient symptoms or complications have been reported as a result of this event.Additional information was received that this was during an olif (oblique lumbar interbody fusion).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTERALIGN¿ SPINAL SYSTEM WITH TITAN NANOLOCK¿ SURFACE TECHNOLOGY
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key23540794
Report Number3003120897-2025-00561
Device Sequence Number9531082
Product Code MAX
UDI-Device Identifier00763000204334
UDI-Public(01)00763000204334(11)230227(10)EM23A032
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K212524
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNAV4680003
Device Catalogue NumberNAV4680003
Device Lot NumberEM23A032
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2025
Initial Report FDA Received Date11/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-