|
On 07 november 2025, it was reported by a sales representative via email that an ar-2324kpsp, swivelock sp peek kl 4.75 mm was reported to have broken during use.This occurred on (b)(6) 2025 during an unknown procedure.It was reported that the peek tip did not pierce through cortical bone: instead appears blunted.Eyelet came away from insertion handle.The case was completed successfully using another anchor.There was case involvement.
|
|
Additional information: g3, h3, h6.Complaint allegation is confirmed by the attached picture, which shows the base of the eyelet broken off and detached from the distal end of the shaft.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The method of bone preparation and the quality of the bone encountered were not specified.Per dfu-0087-eo rev 3 - g precautions - 3.Make sure to use the recommended drill bit or punch to create the bone socket.4.Pushlock and swivelock suture anchors only: during anchor insertion, ensure that the angle of anchor insertion is coaxial to that of the previously prepared bone socket.5.Pushlock and swivelock suture anchors only: insert the driver into the bone socket until the anchor body makes contact with the bone.Preview and adjust suture tension, if necessary.Tension will not increase during final advancement of the anchor body.6.Pushlock and swivelock suture anchors only: ensure that the anchor body is in full contact with the bone before advancing the anchor body into the prepared bone socket.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, misalignment insertion, and/or prying/leveraging of the device during insertion.
|