| Model Number |
500001002500 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaints with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 06-dec-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that station had reboot loop issue.A field service engineer (fse) replaced the power supply and tested all drawers, the remote manager (rm), and the power and battery in the hardware test application.Fse performed preventative maintenance.The system functioned as intended after the field service engineer repaired the device.
|
| |
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ es station on reboot loop.The customer stated that this malfunction occurred while dispensing the medication, and the user managed to get the medication from another station, which caused a delay to the patient care.There were no adverse events or injuries reported based on this event.
|
| |
|