| Catalog Number |
A12MA02102012 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a k-file m-access 21mm 020 file broke during use.The broken part was retrieved by microscope and ultrasound.Patient outcome is unknown as of this mdr.Further information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1884380).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique (hand used file - technique no more verifiable to date), overuse (number of uses not communicated), excessive wear, patient condition and behavior during treatment or material issue (no analysis of the broken fragments possible).In addition, there are further circumstances on the conditions in dentistry (used disinfectants, training/knowledge status), or any other environmental conditions, which are unknown to us and may also have an impact on the reported failure mode.Additional information: we received the following information regarding this complaint: 1.The patient was not injured and did not require treatment.It was the doctor's excessive force during the operation that caused the instrument to separate.2.The product has been discarded and cannot be returned.This is to correct and remove the codes that were initially reported - removing codes for: health effect - clinical code - 4580 health effect - impact code - 4648 the correct codes for this complaint are: health effect - clinical code - 4582 health effect - impact code - 2199 this is a follow up report to correct this code.
|
| |
|
Search Alerts/Recalls
|