Model Number ARDPWT239246A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Getinge became aware of an issue with one of surgical lights ¿ powerled ii.As it was stated, the cover at the base of the spring arm is slightly detached, causing the brake to be loose.We decided to report the issue in abundance of caution as the issue can lead to contamination of sterile field.Based on the information collected, it was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.It is unknown if claimed device was or was not being used for patient treatment or diagnosis when the event took place.According to analysis provided by subject matter expert at manufacturer site, the incident is due to an inappropriate use.The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.Additionally, the users are requested to pay attention to cracks in plastic parts.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Event or Problem Description
Getinge became aware of an issue with one of surgical lights ¿ powerled ii.As it was stated, the cover at the base of the spring arm was slightly detached, causing the brake to be loose.We decided to report the issue in abundance of caution as the issue can lead to contamination of sterile field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET POWERLED II
Common Device Name
LAMP, SURGICAL
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key23542607
Report Number9710055-2025-0000288
Device Sequence Number9683179
Product Code FTD
UDI-Device Identifier03700712412975
UDI-Public(01)03700712412975
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberARDPWT239246A
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2025
Initial Report FDA Received Date11/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-