| Catalog Number |
81324 |
| Medical Device Problem Code |
Restricted Flow rate (1248)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
While using a cavitron select sps g124, they allege that they have low water flow and the handpiece gets hot; no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
Root cause is unknown as no product was returned by the customer.This complaint will be closed but will be reopened if product is returned for evaluation.
|
| |
|
Search Alerts/Recalls
|