This event has been recorded by zimmer biomet under (b)(4).The following sections have been corrected/updated: b4, b5, d2, d4, d9, g1, g3, g6, h1, h2, h3, h4, h6, and h11.Review of the most recent repair record determined the device passed the sample mesh test during evaluation; however, the cutter was damaged, the bottom roller was discolored, and the device was out of calibration.The cutter and bottom roller were damaged, and the device was recalibrated to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|