| Catalog Number |
442022 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd bactec¿ plus anaerobic/f culture vials (plastic) there was one false negative result.The matrix was not blood.No adverse impact was reported regarding the patient or user.This is report 74 of 90.
|
| |
|
Additional Manufacturer Narrative
|
|
E.1: initial reporter addr 1: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd bactec¿ plus anaerobic/f culture vials (plastic) there was one false negative result.The matrix was not blood.No adverse impact was reported regarding the patient or user.This is report 74 of 90.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: catalog 442022.Batch no.5121637.Customer reported a false negative result.Neither photos nor returned good samples were received.Bd was unable to reproduce customer¿s experience with bactec product.Satisfactory results were obtained from retention samples when tested for microbial instrument detection as per quality procedures.Batch history record review did not identify any evidence for which the customer submitted the complaint (i.E.False negative).Microbial instrument detection indicated that the product is performing as expected.A complaint history review was conducted and only the current complaint was found relating to the incident lot number and the ¿as reported¿ defect code.Complaint is unconfirmed based on retention samples and batch record review results.Product insert warnings and precautions sections states that prior to use, each vial should be examined for evidence of contamination such as cloudiness, bulging or depresses septum, or leakage.Vials showing evidence of contamination should not be used.Also, prior to use, the user should examine the vial for evidence of damage or deterioration.Vials displaying turbidity, contamination, or discoloration (darkening) should not be used.The specimen must be collected using sterile techniques to reduce the chance of contamination.No corrective actions were required.A cross-functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigational process.
|
| |
|
Search Alerts/Recalls
|