BOSTON SCIENTIFIC CORPORATION INOGEN? EL ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number D143
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited atrial tachy response (atr) events due to farfield oversensing.This icd remains in service.No adverse patient effects were reported.Additional information received detailed that this icd exhibited oversensing of the right atrial (ra) lead signals.Technical services (ts) was consulted and recommended reprogramming options.This icd remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN? EL ICD DR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23543653
Report Number2124215-2025-82651
Device Sequence Number13205499
Product Code LWS
UDI-Device Identifier00802526534263
UDI-Public(01)00802526534263(17)161218
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/18/2016
Device Model NumberD143
Device Catalogue NumberD143
Device Lot Number197531
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2025
Initial Report FDA Received Date11/13/2025
Date Device Manufactured12/18/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexMale
-
-