| Catalog Number |
2112341 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
According to the complainant, fibers were found inside bag.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).One (1) picture was provided for evaluation.Upon a visual inspection of the image provided, the presence of a ball of lint was observed, in correspondence with one of the pumps.It was determined that the fibers are on the external surface of the tubing; thus, external to the fluid path.Based on the investigation results, the reported defect was confirmed.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.A batch record review for the reported lot number was performed, and no non-conformances or deviations were identified during the manufacturing process or final inspections.The retained sample corresponding to the reported lot number was visually and physically evaluated in accordance with the applicable specifications.All test results were found to be acceptable.As a result, the defect reported by the customer could not be confirmed in the retained sample.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|