MAUDE Adverse Event Report: STAAR SURGICAL EVO/EVO+VISIAN IMPLANTABLE COLLAMER LENS; PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL EVO/EVO+VISIAN IMPLANTABLE COLLAMER LENS; PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5 13.2 (-5.0 / 1.0)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 09/30/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6- health effect- clinical code: 4581- significant reduction of iridocorneal angles.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #: (b)(4).
 
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, significant reduction of iridocorneal angles, and shallowing of the ac.The lens was exchanged for a lens of the same length.The problem was resolved.Cause of the event was reported as unknown.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
 
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, significant reduction of iridocorneal angles, and shallowing of the ac.The lens was exchanged for a shorter length lens and the problem was resolved.Cause of the event was reported as unknown.There are multiple medical device reports associated with this patient.This report is 2 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
 
Additional Manufacturer Narrative
Additional information: b5 - describe event or problem: a facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, significant reduction of iridocorneal angles, and shallowing of the ac.The lens was exchanged for a shorter length lens and the problem was resolved.Cause of the event was reported as unknown.There are multiple medical device reports associated with this patient.This report is 2 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.D4 - model #: vticm5 13.2 (-45.0 / 1.0) catalog #: vticm5 13.2 serial #:(b)(6).Expiration date: 31-mar-2027 udi: (b)(4).H4 - device manufacture date: 26-apr-2024 claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO+VISIAN IMPLANTABLE COLLAMER LENS
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key23548310
Report Number2023826-2025-01725
Device Sequence Number13191495
Product Code QCB
UDI-Device Identifier00840311311088
UDI-Public(01)00840311311088(11)240426(17)270331
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICM5 13.2 (-5.0 / 1.0)
Device Catalogue NumberVTICM5 13.2
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2025
Supplement Date Received by Manufacturer11/18/2025
Initial Report FDA Received Date11/13/2025
Supplement Report FDA Received Date12/11/2025
Date Device Manufactured04/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL - SFC-45, LOT#: UNK.; FOAM TIP PLUNGER - FTP, LOT#: UNK.; INJECTOR MODEL - MSI-PF, LOT#: UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
Patient SexFemale
-
-