• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER Back to Search Results
Model Number MR850JHU
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Method: the subject mr850 respiratory humidifier was returned to the fisher & paykel (f&p) healthcare's regional service centre where it was evaluated by a trained technician.Results: testing of the respiratory humidifier confirmed that no sound was generated from the speaker.Conclusion: the cause of the speaker failure was not determined.Our mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 respiratory humidifier.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 is equipped with visual alarm indicators in addition to the audible alarm.
 
Event or Problem Description
A healthcare facility in (b)(6) reported a fault with an mr850 respiratory humidifier.During service, it was identified that the audible alarm was not functioning.There was no patient involvement reported.
 
Event or Problem Description
A healthcare facility in kansas reported a fault with an mr850 respiratory humidifier.During service, it was identified that the audible alarm was not functioning.There was no patient involvement reported.
 
Additional Manufacturer Narrative
(b)(6).Method: the pcb of the subject mr850 respiratory humidifier was returned to fisher & paykel (f&p) healthcare in new zealand for evaluation.Results: testing of the subject pcb confirmed that no sound was generated from the speaker.Additional resistance testing identified an open circuit in the speaker's coil.Conclusion: the cause of the speaker failure was an open circuit in the speaker coil.The mr850 respiratory humidifier product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 respiratory humidifier.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 respiratory humidifier is equipped with visual alarm indicators in addition to the audible alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
RESPIRATORY HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key23550393
Report Number9611451-2025-01415
Device Sequence Number15882671
Product Code BTT
UDI-Device Identifier09420012407290
UDI-Public(01)09420012407290(10)2101902533(11)211117
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 02/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR850JHU
Device Catalogue NumberMR850JHU
Device Lot Number2101902533
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2025
Supplement Date Received by Manufacturer01/23/2026
Initial Report FDA Received Date11/13/2025
Supplement Report FDA Received Date02/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-