| Catalog Number |
81250170S |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Fall (1848); Nonunion/Delayed-union Bone Fracture (2369)
|
| Date of Event |
10/17/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Upon completion of the investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that a gamma nail broke distally and the patient had to go in for a revision but it was a non union and he had fallen again.
|
| |
|
Event or Problem Description
|
|
It was reported that a gamma nail broke distally and the patient had to go in for a revision but it was a non union and he had fallen again.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction - please refer to d9 - product not returned.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|