BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. INSYTE-N AUTOGUARD; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 381511
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code Abrasion (1689)
Date of Event 10/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event or Problem Description
When this product was inserted into patients, they experienced their skin being ¿shredded.¿ upon visualization of the tip of the catheters, the tips were frayed.This occurred with several attempts using catheters from the same lot.
 
Event or Problem Description
No new information.
 
Additional Manufacturer Narrative
A device history record review was completed by our quality engineer team for provided material number 381911 and lot number 5210670.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported issue, and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSYTE-N AUTOGUARD
Common Device Name
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key23551070
Report Number1710034-2025-01763
Device Sequence Number9609764
Product Code FOZ
UDI-Device Identifier00382903819119
UDI-Public(01)00382903819119(10)5210670
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K201075
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number381511
Device Lot Number5210670
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2025
Supplement Date Received by Manufacturer11/18/2025
Initial Report FDA Received Date11/13/2025
Supplement Report FDA Received Date11/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-