| Catalog Number |
381511 |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
Abrasion (1689)
|
| Date of Event |
10/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
When this product was inserted into patients, they experienced their skin being ¿shredded.¿ upon visualization of the tip of the catheters, the tips were frayed.This occurred with several attempts using catheters from the same lot.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record review was completed by our quality engineer team for provided material number 381911 and lot number 5210670.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported issue, and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.
|
| |
|
Search Alerts/Recalls
|