| Model Number |
KEX152EB |
| Medical Device Problem Codes |
Product Quality Problem (1506); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis : part # kex152eb : lot # 231122650 visual and optical inspection confirmed the connector tip of the syringe is broken and bent due to overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having balloon kyphopl asty (l2) due to primary osteoporosis (compression fracture).It was reported that the connector on the ibt side and the connection to the balloon were deformed.The same thing happened when be was used previously, so it is considered a problem with the quality of the product.There were no patient symptoms reported.There were no further complications reported regarding the event.
|
| |
|