• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. INGENIA; FULL-BODY MRI SYSTEM, SUPERCONDUCTING MAGNET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. INGENIA; FULL-BODY MRI SYSTEM, SUPERCONDUCTING MAGNET Back to Search Results
Model Number INGENIA 3.0T CX
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
Philips received a report that the quadrature body coil (qbc) seal was loose.This seal is used for protection, to avoid the patient from potentially contacting the sharp edges of the magnet covers.If this gasket came out, it is likely that this could lead to serious injury.
 
Additional Manufacturer Narrative
The qbc seal was loose that was replaced and the system was handed back to the customer for use.The failure of the qbc seal to remain in place is considered a malfunction.Remedial action: a field action was initiated under 2023-pd-mr-013.A field safety notification was sent to the customers informing them about this potential issue.The field correction has since been made available to the field in dec 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENIA
Common Device Name
FULL-BODY MRI SYSTEM, SUPERCONDUCTING MAGNET
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs street
cambridge, MA 02141
MDR Report Key23551964
Report Number3003768277-2025-014594
Device Sequence Number15882561
Product Code LNH
UDI-Device Identifier00884838098312
UDI-Public(01)00884838098312
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K173356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Modification/Adjustment
Type of Report Initial
Report Date (Section B) 11/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberINGENIA 3.0T CX
Device Catalogue Number782105
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2025
Initial Report FDA Received Date11/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2023-PD-MR-013
Patient Sequence Number1
Patient SexUnknown
-
-