• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING NETHERLANDS TECNIS SIMPLICITY; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING NETHERLANDS TECNIS SIMPLICITY; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DRN00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); Halo (2227)
Date of Event 10/15/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section h3: the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the patient was unable to tolerate glare and halos after the implantation of the preloaded multifocal intraocular lenses (iols) in both eyes.The lenses were subsequently explanted and replaced with non-johnson and johnson iols.It was reported that the patient had difficulty performing one or more daily activities following the surgery.The patient was dissatisfied with the lenses, which were noted to have caused or contributed to the event.It is unknown if the patient experienced a loss of two or more lines of best spectacle corrected visual acuity (bscva) or if the patient was under or over corrected.The patient¿s pre-operative and post-operative refraction, bscva, and the intended target refraction were also unknown.There were no pre-existing conditions that affected calculation.No unplanned surgical interventions or medication outside the standard of care was required.It was reported that the patient's outcome was ¿as expected¿.No further information was provided.This report is to capture the right eye event.A separate report will be submitted to capture the left eye event.
 
Additional Manufacturer Narrative
The product was received for evaluation.Additional information: section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: nov 20, 2025.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification was performed on the suspect product and found the lens was cut in half and no further issues were identified.No further evaluation was performed.No issues that could cause or contribute to the complaint issues reported were identified during product evaluation.The ¿lens cut¿ observed during the product evaluation could not be confirmed to be related to the manufacturing or design process.Conclusion: based on the complaint investigation results, the product was released within specifications.No product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key23554286
Report Number3012236936-2025-000300
Device Sequence Number21345398
Product Code MFK
UDI-Device Identifier05050474810808
UDI-Public(01)05050474810808(17)271212
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDRN00V
Device Catalogue NumberDRN00V0220
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2025
Supplement Date Received by Manufacturer12/03/2025
Initial Report FDA Received Date11/14/2025
Supplement Report FDA Received Date12/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexFemale
Patient Weight64 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-