| Catalog Number |
362035 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the user facility: details of complaint (reported issue): educator filling the report.Patient transferred from the operating room to the unit alarm identified when patient settled into a room.Nurse attempted doing all of the tips to resolve air in the line.Air alarm would continue to alarm.Nurse attempted resolving more than 5 times with the same outcome.Eventually changed the tubing and the issue resolved.Impact to patient: interruptions to clinical care due to time required to resolve alarms.Action(s) taken changed out pump (isolate and send to bioengineering department), followed b.Braun tips to resolve alarms (air-in-line and occlusion), repeated priming to clear air, tubing replaced (triple bag and send to material management).Medication / fluid ringers lactate.Iv rate 100 ml/hr.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation, the root cause of the reported incident was unable to be determined.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|