Catalog Number PR-05052-HP
Medical Device Problem Codes Obstruction of Flow (2423); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "the head performs the puncture and inserts the guide, but obstruction is present despite raising the arm and removing the tourniquet.The guide is removed and found to be bent, therefore, insertion is performed in the antecubital area of the msd, the puncture is successful, an accessory guide is inserted, and when the guide is returned, no outlet is obtained through the picc lumen, so the picc is returned and the intensive care physician is notified for cvc." there was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The actual device was not returned; however, the customer provided two photos for analysis.The complaint of a kinked guidewire was able to be confirmed by the photos.One photo shows the guidewire in a clinical setting with multiple kinks and bends in the body.A complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions-for-use (ifu) provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested." without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "the head performs the puncture and inserts the guide, but obstruction is present despite raising the arm and removing the tourniquet.The guide is removed and found to be bent, therefore, insertion is performed in the antecubital area of the msd, the puncture is successful, an accessory guide is inserted, and when the guide is returned, no outlet is obtained through the picc lumen, so the picc is returned and the intensive care physician is notified for cvc." there was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".Associated mdr numbers include: 9680794-2025-00936 and 9680794-2025-00930.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW PICC SET 2-LUMEN: 5 FR X 19-5/8IN (50 CM
Common Device Name
CATHETER, INTRAVASCULAR, THER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key23555011
Report Number9680794-2025-00930
Device Sequence Number10053060
Product Code LJS
UDI-Device Identifier10801902095388
UDI-Public(01)10801902095388(11)250226(17)270131(10)14F25A0270(21)
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K042126
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberPR-05052-HP
Device Lot Number14F25A0270
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2025
Supplement Date Received by Manufacturer12/10/2025
Initial Report FDA Received Date11/14/2025
Supplement Report FDA Received Date12/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-