| Catalog Number |
PWKIT03 |
| Medical Device Problem Code |
Decrease in Suction (1146)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the purewick accessory replacement kit failed suction test.We have requested that the customer work directly with the manufacturer first to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states, note: the purewick¿ urine collection system should be stationary while in use.Device is not designed to be used while in transport.5.Place the collection canister (c) in the purewick¿ urine collection system base and press down firmly on the lid making sure the lid is sealed.Optional (prior to sealing canister lid): while operating the purewick¿ urine collection system, ammonia odor for up to 1800 ml of urine can be eliminated by adding 2 teaspoons (10 grams) of vitamin c (ascorbic acid) powder into the collection canister before use.At least 99.93% of pure vitamin c (ascorbic acid) is recommended.If using the privacy cover (j), slip privacy cover onto canister (similar to a coffee cup sleeve) prior to placing canister into the base.Attach the pump tubing (d) to the purewick¿ urine collection system connector port (f) and the connector port (e) on the collection canister troubleshooting steps for " the device is on but is not suctioning properly": 1.Ensure tubing connections are connected properly.2.Check collector tubing for blockage or flow restriction such as pinched or kinked tubing.3.Ensure overflow stop valve in collection canister lid is open.The valve floats to the top when the collection canister is full.The stop valve may close if the lid or canister is tipped sideways or upside down.Disconnect tubing and gently shake the lid to reset the valve down to the open position.4.Ensure collection canister is sealed with the lid tightly closed.5.Verify suction by disconnecting purewick¿ external catheter from the collector tubing and placing the end of the collector tubing into a cup of water.If water easily flows into the collection canister replace the purewick¿ external catheter.A dhr could not be performed since no lot number was provided.No additional actions can be taken at this time.Correction: d,e,h udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the purewick accessory replacement kit failed suction test.We have requested that the customer work directly with the manufacturer first to resolve the issue.
|
| |
|