C.R. BARD INC. (COVINGTON) -1018233 PUREWICK URINE COLLECTION SYSTEM Back to Search Results
Catalog Number PWKIT03
Medical Device Problem Code Decrease in Suction (1146)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the purewick accessory replacement kit failed suction test.We have requested that the customer work directly with the manufacturer first to resolve the issue.
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states, note: the purewick¿ urine collection system should be stationary while in use.Device is not designed to be used while in transport.5.Place the collection canister (c) in the purewick¿ urine collection system base and press down firmly on the lid making sure the lid is sealed.Optional (prior to sealing canister lid): while operating the purewick¿ urine collection system, ammonia odor for up to 1800 ml of urine can be eliminated by adding 2 teaspoons (10 grams) of vitamin c (ascorbic acid) powder into the collection canister before use.At least 99.93% of pure vitamin c (ascorbic acid) is recommended.If using the privacy cover (j), slip privacy cover onto canister (similar to a coffee cup sleeve) prior to placing canister into the base.Attach the pump tubing (d) to the purewick¿ urine collection system connector port (f) and the connector port (e) on the collection canister troubleshooting steps for " the device is on but is not suctioning properly": 1.Ensure tubing connections are connected properly.2.Check collector tubing for blockage or flow restriction such as pinched or kinked tubing.3.Ensure overflow stop valve in collection canister lid is open.The valve floats to the top when the collection canister is full.The stop valve may close if the lid or canister is tipped sideways or upside down.Disconnect tubing and gently shake the lid to reset the valve down to the open position.4.Ensure collection canister is sealed with the lid tightly closed.5.Verify suction by disconnecting purewick¿ external catheter from the collector tubing and placing the end of the collector tubing into a cup of water.If water easily flows into the collection canister replace the purewick¿ external catheter.A dhr could not be performed since no lot number was provided.No additional actions can be taken at this time.Correction: d,e,h udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the purewick accessory replacement kit failed suction test.We have requested that the customer work directly with the manufacturer first to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK URINE COLLECTION SYSTEM
Common Device Name
PUREWICK
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23555286
Report Number1018233-2025-10282
Device Sequence Number10132831
Product Code NZU
UDI-Device Identifier00801741185380
UDI-Public(01)00801741185380
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberPWKIT03
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2025
Supplement Date Received by Manufacturer01/15/2026
Initial Report FDA Received Date11/14/2025
Supplement Report FDA Received Date01/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-