MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ORTHOPEDIC SALVAGE SYSTEM POLYETHYLENE TIBIAL BUSHING; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ORTHOPEDIC SALVAGE SYSTEM POLYETHYLENE TIBIAL BUSHING; PROSTHESIS, KNEE Back to Search Results
Catalog Number 150476
Medical Device Problem Codes Unstable (1667); Patient-Device Incompatibility (2682); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Hyperextension (4523)
Date of Event 10/20/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a left knee arthroplasty revision to address excessive hyperextension approximately one (1) year post-operatively.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).D10: concomitant devices - orthopedic salvage system rs segmental femoral component left 8.5cm catalog#: 161124, lot#: 66097829, orthopedic salvage system rs tibial bearing 14mm catalog#: 161095, lot#: 65980346, orthopedic salvage system reinforced yoke catalog#: 150493, lot#: 66294444, orthopedic salvage system rs axle catalog#: 161035, lot#: 66654338, orthopedic salvage system poly locking pin catalog#: 150478, lot#: 66573572, biomet series a standard 3-peg patella 31mm catalog#: 184764, lot#: 66339705.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d9, g3, g6, h2, h6, h10, h11.
 
Event or Problem Description
It was reported that the patient underwent a left knee arthroplasty revision to address excessive hyperextension and instability approximately one (1) year post-operatively.During the revision, the femoral bushings and polyethylene tibial bearing were identified to have fractured.No additional patient consequences were reported.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4, g3, g6, h2, h3, h6, h10, h11.The following sections were corrected: h4.The device was evaluated, however, the reported event could not be confirmed.Visual evaluation of the returned device identified signs of use and was dinged/nicked on the wide end.Dimensional analysis of the product determined it was conforming to print specifications where measured.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORTHOPEDIC SALVAGE SYSTEM POLYETHYLENE TIBIAL BUSHING
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23555711
Report Number0001825034-2025-03621
Device Sequence Number10053079
Product Code JDI
UDI-Device Identifier00880304002258
UDI-Public(01)00880304002258(17)290826(10)66873956
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K002757
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number150476
Device Lot Number66873956
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2025
Supplement Date Received by Manufacturer11/14/2025
03/24/2026
Initial Report FDA Received Date11/14/2025
Supplement Report FDA Received Date12/09/2025
03/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age80 YR
Patient SexFemale
-
-