A healthcare professional reported that during a cataract and intraocular lens (iol) implant procedure, while injecting, the plunger overrode the lens.There was no issue with the leading or trailing haptic and there was patient contact.No medical intervention was required and the patient was not hospitalized as a result of this event.The device was maintained between 18 degrees celsius (64 degrees fahrenheit) and 23 degrees celsius (73 degrees fahrenheit) at operating room temperatures after being allowed to acclimate for at least 30 minutes.The procedure was completed on the same day using a different lens without any harm to the patient.Additional information was requested.
|
|
Additional information was provided in d.9., h.3., h.6 and h.11.The used device and the lens were returned in the blister tray inside the carton.The plunger lock and lens stop were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was retracted to the starting point of the loading area.The lens was advanced to mid nozzle.The leading haptic was extended.The trailing haptic was folded onto the anterior optic surface.The nozzle was removed and cleaned for further evaluation.The lens was removed during cleaning.Top coat dye stain testing was conducted with acceptable results.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A non-qualified viscoelastic was used.The root cause for the reported complaint may be related to a failure to follow the instructions for use (ifu).A non-qualified viscoelastic was indicated.The ifu instructs: during device preparation and implantation of the company intra ocular lens (iol) with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.The ifu instructs: take at least 7 seconds to gently fold the intraocular lens (iol) by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the fill-to line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.The ifu instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be out of position can result in a broken haptic or other negative outcome, since the haptic may be trapped and stretched, and/or pinched and sheared by the moving plunger.The position of the plunger in relation to the lens during the delivery attempt cannot be determined.The plunger was retracted upon return.The evaluation of this used sample cannot confirm if the reported plunger override had occurred.Plunger override may occur: due to rapid advancement faster than the ifu recommended rate.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (greater than 23 degrees centigrade / 73 degrees fahrenheit) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Top coat dye stain testing was conducted with acceptable results.The manufacturer internal reference number is: (b)(4).
|