| Catalog Number |
07027133190 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys cmv igm assay results for a patient sample tested on the cobas e 801 module.On (b)(6) 2025: elecsys cmv igm result was 0.58 coi (non-reactive).Elecsys cmv igg result was 24.4 u/ml (reactive).On (b)(6) 2025: vidas cmv igg result was 25 au/ml (reactive).Vidas cmv igg avidity result was 40 avi% (equivocal).Diasorin liaison cmv igm result was 32.7 au/ml (reactive).Diasorin liaison cmv igg result was 69.8 au/ml (reactive) diasorin liaison cmv igg avidity result was 0.199.On (b)(6) 2025: elecsys cmv igm result was 0.54 coi (non-reactive).Elecsys cmv igg result was 21.3 u/ml (reactive).
|
| |
|
Additional Manufacturer Narrative
|
|
The serial number of the analyzer was (b)(6).The remaining sample volume was sent to another laboratory for additional testing.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Qc and calibration data were found to be within the expected ranges.No further results were provided from the additional testing performed at the other laboratory.Sample material was requested; however, no sample volume was left.Therefore, further investigation could not be performed.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|
Search Alerts/Recalls
|